Modes of mechanical ventilation

Modes of mechanical ventilation are one of the most important aspects of the usage of mechanical ventilation. The mode refers to the method of inspiratory support. Mode selection is generally based on clinician familiarity and institutional preferences since there is a paucity of evidence indicating that the mode affects clinical outcome. The most frequently used forms of volume-limited mechanical ventilation are IMV and CMV.[1] Substantial changes in the nomenclature of mechanical ventilation over the years but more recently has become standardized by many respirology/pulmonology groups.[2][3]

Contents

Negative pressure ventilation

Negative pressure ventilation works by producing an intermittent negative pressure around the chest and abdomen. Negative pressure moves across the chest and diaphragm and causes air to move into the lungs in the normal fashion.[4] When the negative pressure stops being applied, the chest returns to atmospheric pressure and the inspired air then is exhaled.[4]

Disadvantages

Negative pressure ventilation has several disadvantages compared to positive-pressure ventilation, such as; The machines are less portable, more difficult to apply, and is infrequently used. Moreover, negative pressure is actually contraindicated in obstructive sleep apnea.[4]

Types of negative pressure ventilators

Positive pressure ventilation

Positive pressure ventilation is achieved by applying positive pressure higher than atmospheric pressure at the airway opening. Increasing the pressure at the airway opening produces a pressure gradient that generates an inspiratory flow. This flow in turn results in the delivery of a breath.

Continuous mandatory ventilation

CMV — Continuous mandatory ventilation (formerly known as Assist Control or AC) is a mode of ventilation where breaths are delivered based on set variables. The patient may initate breaths by attempting to breathe. Once a breath is initated, either by the patient or by the ventilator the set tidal volume is delivered. Continuous mandatory ventilation used to also be called Volume Control or Assist Control Volume Control (AC/VC), though this is no longer recommended. Since nomenclature of mechanical ventilation is only recently standardized[5] there are many different names that historically were used to reference CMV but now reference Assist Control.[5] Names such as: volume control ventilation, and volume cycled ventilation in modern usage refer to the Assist Control mode.

Controlled mechanical ventilation vs. Continuous mandatory ventilation

Controlled mechanical ventilation in its original form had no patient sensitivity. A breath set was a breath delivered. Continuous mandatory ventilation was created out of the need for patient-initiation in breaths. Fundamentally, Continuous mandatory ventilation is controlled mechanical ventilation (CMV) with a sensitivity for patient breathing. The use of controlled mechanical ventilation requires the patient be completely unconscious, either pharmacokinetically or otherwise in a coma.

Expected outcomes and considerations

Continuous mandatory ventilation (formerly Assist Control or AC) is associated with profound diaphragm muscle dysfunction and atrophy.[6] Continuous mandatory ventilation is no longer the preferred mode of mechanical ventilation.[7]

Intermittent mandatory ventilation

IMV — Intermittent mandatory ventilation is similar to continuous mandatory ventilation in two ways: the minute ventilation (VE) is determined (by setting the respiratory rate and tidal volume); and the patient is able to increase the minute ventilation. However, IMV differs from continuous mandatory ventilation in the way that the minute ventilation is increased. Specifically, patients increase the minute ventilation by spontaneous breathing, rather than patient-initiated ventilator breaths. The ventilator breaths are synchronized with patient inspiratory effort.[8][9] IMV with pressure support is the most efficient and effective mode of mechanical ventilation.[10] Intermittent mandatory ventilation has not always had the synchronized feature, so the division of modes were understood to be SIMV (synchronized) vs IMV (not-synchronized). Since the American Association for Respiratory Care established a nomenclature of mechanical ventilation the "synchronized" part of the title has been dropped and now there is only IMV.

Volume-control

VC — Volume-controlled ventilation (formerly called volume-limited or volume-cycled ventilation) requires the clinician to set the peak flow rate, flow pattern, tidal volume, respiratory rate, positive end-expiratory pressure (applied PEEP), and fraction of inspired oxygen (FiO2). Inspiration ends after delivery of the set tidal volume.

Mandatory minute ventilation (MMV)

MMV — Mandatory minute ventilation allows spontaneous breathing with automatic adjustments of mandatory ventilation to the meet the patient’s preset minimum minute volume requirement. If the patient maintains the minute volume settings for VT x f, no mandatory breaths are delivered. If the patient's minute volume is insufficient, mandatory delivery of the preset tidal volume will occur until the minute volume is achieved. The method for monitoring whether or not the patient is meeting the required minute ventilation (VE) differs by ventilator brand and model, but generally there is a window of monitored time, and a smaller window checked against the larger window (i.e., in the Dräger Evita® line of mechanical ventilators there is a moving 20-second window, and every 7 seconds the current tidal volume and rate are measured) to decide whether a mechanical breath is needed to maintain the minute ventilation. MMV is the optimal mode for weaning in neonatal and pediatric populations and has been shown to reduce long term complications related to mechanical ventilation.[11]

Pressure regulated volume control (PRVC)

PRVC — Pressure regulated volume control is an IMV based mode. Pressure regulated volume control utilizes pressure-limited, volume-targeted, time-cycled breaths which can be either ventilator or patient initiated. The peak inspiratory pressure delivered by the ventilator is varied on a breath-to-breath basis to achieve a target tidal volume which is set by the clinician. For example, if a target tidal volume of 500 mL is set but the ventilator delivers 600 mL, the next breath will be delivered with a lower inspiratory pressure to achieve a lower tidal volume. Though PRVC is regarded as a hybrid mode because of its tidal-volume (VC) settings and pressure-limiting (PC) settings fundamentally PRVC is a volume-control mode.

Pressure-control

PC — Pressure Control (PC) is a controlled mode of ventilation. The ventilator delivers a flow to maintain the preset pressure at a preset respiratory rate over a preset inspiratory time.[12] The pressure is constant during the inspiratory time and the flow is decelerating. If for any reason pressure decreases during inspiration, the flow from the ventilator will immediately increase to maintain the set inspiratory pressure.[13]

Continuous spontaneous ventilation

Airway pressure release ventilation (APRV)

APRV — Airway Pressure Release Ventilation is a time-cycled alternant between two levels of positive airway pressure, with the main time on the high level and a brief expiratory release to facilitate ventilation.[14] This is a type of inverse ratio ventilation. The exhalation time (Tlow) is shortened to usually less than one second to maintain alveoli inflation. Fundamentally this is a continuous pressure with a brief release. APRV currently the most efficient conventional mode for lung protective ventilation.[15]

Different perceptions of this mode may exist around the globe. While 'APRV' is common to users in North America, a very similar mode, biphasic positive airway pressure (BIPAP), was introduced in Europe.[16] The term APRV has also been used in American journals where, from the ventilation characteristics, BIPAP would have been the appropriate terminology.[17] To further confusion, BiPAP© is a registered trade-mark for a noninvasive ventilation mode in a specific ventilator (Respironics Inc.). Other names (BILEVEL, DUOPAP, BIVENT) have been created for legal reasons. Although similar in modality, these terms describe how a mode is intended to inflate the lung, rather than defining the characteristics of synchronization or the way spontaneous breathing efforts are supported.

Continuous positive airway pressure (CPAP)

CPAP — Continuous positive airway pressure (CPAP) refers to the delivery of a continuous level of positive airway pressure. It is functionally similar to PEEP. The ventilator does not cycle during CPAP, no additional pressure above the level of CPAP is provided, and patients must initiate all breaths. Nasal CPAP is frequently used in neonates though its use is controversial. Studies have shown nasal CPAP to reduce ventilator time but an increased occurrence of pneumothorax was also prevalent.[18]

Bilevel positive airway pressure (BPAP)

BPAP — Bilevel positive airway pressure (BPAP) is a mode used during noninvasive positive pressure ventilation (NPPV). First used in 1988 by Professor Benzer in Germany,[19] it delivers a preset inspiratory positive airway pressure (IPAP) and expiratory positive airway pressure (EPAP). BPAP can be described as a Continuous Positive Airway Pressure system with a time-cycled change of the applied CPAP level.[20] CPAP, BPAP and other non-invasive ventilation modes have been shown to be effective management tools for chronic obstructive pulmonary disease and acute respiratory failure.[21]

Often BPAP is incorrectly referred to as "BiPAP". BiPAP® is the name of a portable ventilator manufactured by Respironics Corporation; it is just one of many ventilators that can deliver BPAP.

Positive End Expiratory Pressure (PEEP)

PEEPPositive end expiratory pressure is pressure applied upon expiration. PEEP is applied either using a valve that is connected to the expiratory port and set manually or a valve managed internally by a mechanical ventilator. PEEP is simply a pressure that an exhalation has to bypass, effectively causing alveoli to remain open and not fully deflate. This mechanism for maintaining inflated alveoli helps increase partial pressure of oxygen in arterial blood, an increase in PEEP increases the PaO2. [22]

Pressure Support (PS)

PS — Pressure support is a spontaneous mode of ventilation also named Pressure Support Ventilation (PSV). The patient initiates every breath and the ventilator delivers support with the preset pressure value. With support from the ventilator, the patient also regulates their own respiratory rate and their tidal volume.

In Pressure Support, the set inspiratory pressure support level is kept constant and there is a decelerating flow. The patient triggers all breaths. If there is a change in the mechanical properties of the lung/thorax and patient effort, the delivered tidal volume will be affected. The user must then regulate the pressure support level to obtain desired ventilation.[23][24]

Pressure support improves oxygenation,[25] ventilation and decreases work of breathing.

High Frequency Ventilation (HFV)

High frequency ventilation
Intervention
MeSH D006612

High frequency ventilation is a type of mechanical ventilation that employs very high respiratory rates (>150 (Vf) breaths per minute) and very small tidal volumes.[26][27] High frequency ventilation is thought to reduce ventilator-associated lung injury (VALI), especially in the context of ARDS and acute lung injury.[26] This is commonly referred to as lung protective ventilation.[28] There are different flavors of High frequency ventilation.[26] Each type has its own unique advantages and disadvantages. The types of HFV are characterized by the delivery system and the type of exhalation phase (active vs passive). High Frequency Ventilation may be used alone, or in combination with conventional mechanical ventilation. In general, those devices that need conventional mechanical ventilation do not produce the same lung protective effects as those that can operate without tidal breathing. Specifications and capabilities will vary depending on the device manufacturer.

Uses and applications

High-frequency ventilation is often used to treat patients with acute lung injury (ALI) or acute respiratory distress syndrome (ARDS) but the effect of this treatment on clinical outcomes has not been well established.[29]

Problems and concerns

High-frequency ventilation currently has no measurement of tidal volume so without an external device or machine a clinician cannot know if the tidal volumes are appropriate or whether or not the patient is at risk for volutrauma.[30]

Active high-frequency ventilation (HFV-A)

HFV-A — The term active refers to the ventilators forced expiratory system. In a HFV-A scenario, the ventilator uses pressure to apply an inspiratory breath and then applies an opposite pressure to force an expiratory breath. In high-frequency oscillatory ventilation (a ventilator patented by sensormedics corp) the oscillation bellow and piston force positive pressure in and apply negative pressure to force an expiration.[31]

Passive high-frequency ventilation (HFV-P)

HFV-P — The term passive refers to the ventilators non-forced expiratory system. In a HFV-P scenario, the ventilator uses pressure to apply an inspiratory breath and then simply returns to atmospheric pressure to allow for a passive expiration.

Liquid ventilation (LV)

Liquid ventilation (LV) — is a technique of mechanical ventilation in which the lungs are insufflated with an oxygenated perfluorochemical liquid rather than an oxygen-containing gas mixture. The use of perfluorochemicals, rather than nitrogen, as the inert carrier of oxygen and carbon dioxide offers a number of theoretical advantages for the treatment of acute lung injury, including:

Despite its theoretical advantages, efficacy studies have been disappointing and the optimal clinical use of LV has yet to be defined.[32]

Total liquid ventilation (TLV)

Total liquid ventilation — In total liquid ventilation (TLV), the entire lung is filled with an oxygenated PFC liquid, and a liquid tidal volume of PFC is actively pumped into and out of the lungs. A specialized apparatus is required to deliver and remove the relatively dense, viscous PFC tidal volumes, and to extracorporeally oxygenate and remove carbon dioxide from the liquid.[33][34][35]

Partial liquid ventilation (PLV)

Partial liquid ventilation — In partial liquid ventilation (PLV), the lungs are slowly filled with a volume of PFC equivalent or close to the FRC during gas ventilation. The PFC within the lungs is oxygenated and carbon dioxide is removed by means of gas breaths cycling in the lungs by a conventional gas ventilator.[36]

Other modes and types of ventilation

Proportional Assist Ventilation (PAV)

PAV — Proportional assist ventilation is a mode in which the ventilator guarantees the percentage of work regardless of changes in pulmonary compliance and resistance.[37] The ventilator varies the tidal volume and pressure based on the patients work of breathing, the amount it delivers is proportional to the percentage of assistance it is set to give.

See also

References

  1. ^ Esteban A, Anzueto A, Alía I, et al. How is mechanical ventilation employed in the intensive care unit? An international utilization review. Am J Respir Crit Care Med 2000; 161:1450.
  2. ^ Donn SM (2009). "Neonatal ventilators: how do they differ?". J Perinatol 29 Suppl 2: S73-8. doi:10.1038/jp.2009.23. PMID 19399015. http://www.ncbi.nlm.nih.gov/entrez/eutils/elink.fcgi?dbfrom=pubmed&tool=sumsearch.org/cite&retmode=ref&cmd=prlinks&id=19399015. 
  3. ^ Chatburn RL, Volsko TA, Hazy J, Harris LN, Sanders S (2011). "Determining the Basis for a Taxonomy of Mechanical Ventilation.". Respir Care. doi:10.4187/respcare.01327. PMID 22004898. http://www.ncbi.nlm.nih.gov/entrez/eutils/elink.fcgi?dbfrom=pubmed&tool=sumsearch.org/cite&retmode=ref&cmd=prlinks&id=22004898. 
  4. ^ a b c d e Hill NS (1986). "Clinical applications of body ventilators.". Chest 90 (6): 897–905. PMID 3536343. 
  5. ^ a b Chatburn RL. Classification of ventilator modes: update and proposal for implementation. Respir Care 2007; 52:301–323.
  6. ^ Sassoon CS, Zhu E, Caiozzo VJ (2004). "Assist-control mechanical ventilation attenuates ventilator-induced diaphragmatic dysfunction.". Am J Respir Crit Care Med 170 (6): 626–32. doi:10.1164/rccm.200401-042OC. PMID 15201132. http://www.ncbi.nlm.nih.gov/entrez/eutils/elink.fcgi?dbfrom=pubmed&tool=sumsearch.org/cite&retmode=ref&cmd=prlinks&id=15201132. 
  7. ^ Macintyre N (2011). "Counterpoint: Is Pressure Assist-Control Preferred Over Volume Assist-Control Mode for Lung Protective Ventilation in Patients With ARDS? No.". Chest 140 (2): 290–2. doi:10.1378/chest.11-1052. PMID 21813526. http://www.ncbi.nlm.nih.gov/entrez/eutils/elink.fcgi?dbfrom=pubmed&tool=sumsearch.org/cite&retmode=ref&cmd=prlinks&id=21813526. 
  8. ^ Sassoon CS, Del Rosario N, Fei R, et al. Influence of pressure- and flow-triggered synchronous intermittent mandatory ventilation on inspiratory muscle work. Crit Care Med 1994; 22:1933.
  9. ^ Christopher KL, Neff TA, Bowman JL, et al. Demand and continuous flow intermittent mandatory ventilation systems. Chest 1985; 87:625.
  10. ^ D. C. Shelledy, J. L. Rau & L. Thomas-Goodfellow (January–February 1995). "A comparison of the effects of assist-control, SIMV, and SIMV with pressure support on ventilation, oxygen consumption, and ventilatory equivalent". Heart & lung : the journal of critical care 24 (1): 67–75. PMID 7706102. 
  11. ^ Scott O. Guthrie, Chris Lynn, Bonnie J. Lafleur, Steven M. Donn & William F. Walsh (October 2005). "A crossover analysis of mandatory minute ventilation compared to synchronized intermittent mandatory ventilation in neonates". Journal of perinatology : official journal of the California Perinatal Association 25 (10): 643–646. doi:10.1038/sj.jp.7211371. PMID 16079905. 
  12. ^ MAQUET, "Modes of ventilation in SERVO-i, Invasive and Non-invasive, 2008 MAQUET Critical Care AB, Order No 66 14 692
  13. ^ MAQUET, "Modes of ventilation in SERVO-s, Invasive and Non-invasive", 2009 MAQUET Critical Care AB, Order No 66 61 131
  14. ^ Dietrich Henzler (2011). "What on earth is APRV?". Critical care (London, England) 15 (1): 115. doi:10.1186/cc9419. PMID 21345265. 
  15. ^ Adrian A. Maung & Lewis J. Kaplan (July 2011). "Airway pressure release ventilation in acute respiratory distress syndrome". Critical care clinics 27 (3): 501–509. doi:10.1016/j.ccc.2011.05.003. PMID 21742214. 
  16. ^ M. Baum, H. Benzer, C. Putensen, W. Koller & G. Putz (September 1989). "[Biphasic positive airway pressure (BIPAP)--a new form of augmented ventilation]". Der Anaesthesist 38 (9): 452–458. PMID 2686487. 
  17. ^ C. Putensen, S. Zech, H. Wrigge, J. Zinserling, F. Stuber, T. Von Spiegel & N. Mutz (July 2001). "Long-term effects of spontaneous breathing during ventilatory support in patients with acute lung injury". American journal of respiratory and critical care medicine 164 (1): 43–49. PMID 11435237. 
  18. ^ Colin J. Morley, Peter G. Davis, Lex W. Doyle, Luc P. Brion, Jean-Michel Hascoet & John B. Carlin (February 2008). "Nasal CPAP or intubation at birth for very preterm infants". The New England journal of medicine 358 (7): 700–708. doi:10.1056/NEJMoa072788. PMID 18272893. 
  19. ^ Benzer H (1988) Ventilatory support by intermittent changes in PEEP levels. 4th European Congress on Intensive Care Medicine. Baveno-Stresa
  20. ^ C. Hormann, M. Baum, C. Putensen, N. J. Mutz & H. Benzer (January 1994). "Biphasic positive airway pressure (BIPAP)--a new mode of ventilatory support". European journal of anaesthesiology 11 (1): 37–42. PMID 8143712. 
  21. ^ M. A. Levitt (November 2001). "A prospective, randomized trial of BiPAP in severe acute congestive heart failure". The Journal of emergency medicine 21 (4): 363–369. PMID 11728761. 
  22. ^ D. P. Schuster, M. Klain & J. V. Snyder (October 1982). "Comparison of high frequency jet ventilation to conventional ventilation during severe acute respiratory failure in humans". Critical care medicine 10 (10): 625–630. PMID 6749433. 
  23. ^ MAQUET, "Modes of ventilation in SERVO-i, invasive and non-invasive", 2008 MAQUET Critical Care AB, Order No 66 14 692
  24. ^ MAQUET, "Modes of ventilation in SERVO-s, invasive and non-invasive", 2009 MAQUET Critical Care AB, Order No 66 61 131
  25. ^ Spieth PM, Carvalho AR, Güldner A, et al. (April 2011). "Pressure support improves oxygenation and lung protection compared to pressure-controlled ventilation and is further improved by random variation of pressure support". Critical Care Medicine 39 (4): 746–55. doi:10.1097/CCM.0b013e318206bda6. PMID 21263322. http://meta.wkhealth.com/pt/pt-core/template-journal/lwwgateway/media/landingpage.htm?issn=0090-3493&volume=39&issue=4&spage=746. 
  26. ^ a b c Krishnan JA, Brower RG (2000). "High-frequency ventilation for acute lung injury and ARDS". Chest 118 (3): 795–807. doi:10.1378/chest.118.3.795. PMID 10988205.  Free Full Text.
  27. ^ Standiford TJ, Morganroth ML. High-frequency ventilation. Chest 1989; 96:1380.
  28. ^ Bollen CW, Uiterwaal CS, van Vught AJ. Systematic review of determinants of mortality in high frequency oscillatory ventilation in acute respiratory distress syndrome. Crit Care 2006; 10:R34.
  29. ^ Wunsch H, Mapstone J (2004). "High-frequency ventilation versus conventional ventilation for treatment of acute lung injury and acute respiratory distress syndrome.". Cochrane Database Syst Rev (1): CD004085. doi:10.1002/14651858.CD004085.pub2. PMID 14974056. 
  30. ^ Sykes MK (1985). "High frequency ventilation.". Thorax 40 (3): 161–5. PMC PMC460016. PMID 3885469. http://www.pubmedcentral.nih.gov/articlerender.fcgi?tool=pmcentrez&artid=PMC460016. 
  31. ^ Allardet-Servent J (2011) High-frequency oscillatory ventilation in adult patients with acute respiratory distress syndrome: Where do we stand and where should we go? Crit Care Med 39 (12):2761-2. DOI:10.1097/CCM.0b013e31822a5c35 PMID: 22094505
  32. ^ Degraeuwe PL, Vos GD, Blanco CE (1995). "Perfluorochemical liquid ventilation: from the animal laboratory to the intensive care unit.". Int J Artif Organs 18 (10): 674–83. PMID 8647601. 
  33. ^ Norris MK, Fuhrman BP, Leach CL (1994). "Liquid ventilation: it's not science fiction anymore.". AACN Clin Issues Crit Care Nurs 5 (3): 246–54. PMID 7780839. 
  34. ^ Greenspan JS (1996). "Physiology and clinical role of liquid ventilation therapy.". J Perinatol 16 (2 Pt 2 Su): S47-52. PMID 8732549. 
  35. ^ Dirkes S (1996). "Liquid ventilation: new frontiers in the treatment of ARDS.". Crit Care Nurse 16 (3): 53–8. PMID 8852261. 
  36. ^ Cox CA, Wolfson MR, Shaffer TH (1996). "Liquid ventilation: a comprehensive overview.". Neonatal Netw 15 (3): 31–43. PMID 8715647. 
  37. ^ Younes M. Proportional assist ventilation, a new approach to ventilatory support. Theory. Am Rev Respir Dis 1992; 145(1):114-120.